Test Code CIRC Circulating Anticoagulant

Cerner Test Code

CIRC

Acceptable Collection Tube(s) or Specimen

Blue

Specimen Draw Volume

5.4 mL

Processing Instructions

Draw 5.4 ml of blood place in 2 Blue Top tubes. Tube must be filled appropriately to the fill line to ensure accurate results. Centrifuge, separate, centrifuge again and freeze.

Stability

Stability: Plasma: Room temp 4 hrs, Refrigerated 4 hrs, Frozen 1 week; Whole Blood: tubes left capped/unspun are stable for 1 hour at room temperature.

Performed Test Frequency

Everyday

CPT(s)

85610|885611 X 6

Methodology

Clotting

Performing Location

MercyOne Des Moines Laboratory, Hematology

Rejection Criteria

Rejection criteria: Refrigerated plasma.